TM 5-6675-312-14
STEP 1. Center and level theodolite (para. 2-8).
STEP 2. Rotate telescope so that optical sight is on
bottom.
STEP 3. Remove four screws that will secure telescope
level to telescope assembly.
STEP 4. Secure telescope level to telescope assembly
with four screws.
STEP 5. Setup a vertical staff approximately 50 yards
(or 50 meters) from theodolite.
NOTE
"Face right" is defined as, when the
can look into the telescope eyepiece (fig. 2-
1). "Face left" is defined as, when the
can look into the telescope eyepiece.
STEP 6. Place theodolite in "face left".
STEP 7. Using the vertical drive screw (fig. 2-1) set the
vertical circle reading scale (V) to 90 00.0'
as indicated in reading microscope eyepiece.
STEP 8. Viewing through the telescope eyepiece (fig.
2-1) note where the horizontal cross hair cuts
the vertical staff.
STEP 9. Place the theodolite in "face right".
STEP 10. Using the vertical drive screw (fig. 2-1) set
the vertical circle reading scale (V) to 270
00.0' as indicated in the reading microscope
eyepiece.
STEP 11. Viewing through the telescope eyepiece (fig.
2-1) note where horizontal cross hair cuts the
vertical staff.
Figure 3-6. Telescope level installation.
STEP 12. Using the vertical drive screw (fig. 2-1) set
telescope eyepiece cross hair to the mean
(center) of readings noted in steps 7 and 9.
STEP 13. Using adjusting pin, adjust telescope level
until the ends of the split bubble are in
coincidence.
3-7