TM 11-6625-2440-12
Seq
Symptom
Probable Cause
Corrective action
No.
29
Scope indication is out of tole-
Circuit card 2A1A8 faulty ........................... Replace circuit card 2A1A8 (para 4-18).
Through
rance
35
Counter or scope indication is out
Circuit card 2A1A1 faulty ........................... Replace circuit card 2A1A2 (para 4-1 8).
Through
of tolerance.
38
39
Scope indication not as specified
Circuit card 2A1A2 faulty ........................... Replace circuit card 2A1A2 (para 4-18).
40
Scope indication not as specified
Circuit card 2A1A3 faulty ........................... Replace circuit card 2A1A3 (para 4-18).
and
41
42
Scope indication not as specified-
Circuit card 2A1A2 faulty ........................... Replace circuit card 2A1A2 (para 4-18).
43
Scope Indication not as specified
Circuit card 2A1A4 or 2A1A10 faulty.
and
If this fails to correct the trouble, replace
44
46
Scope indication not as specified
Circuit card 2A1A4 faulty ........................... Replace circuit card 2A1A4 (para 4-18)
47
Counter indication out of tole-
Circuit card 2A1A6 or 2A1A8 faulty.......
and
rance
this fails to correct the trouble. replace
48
49
Counter indication out of tole-
Circuit card 2A1A7or 2A1A8 faulty
and
rance
If this fails to correct the trouble, replace
50
51
Counter indication is not specified
Circuit card 2A1A5 or 2A1A8 faulty
and
If this fails to correct the trouble,. replace
52
53
GC(2 indicator does not go out
Circuit card 2A1A1 or 2A1A16 faulty
or GC1 indicator does not light.
If this fails to correct the trouble, replace
54
Counter indication is not as spe-
Circuit card 2A1A15 or 2A1A10 faulty ........ Replace circuit card 2A1A15 (para 4-18) If
cified.
this fails to correct the trouble, replace
55
GC1 indicator does not go out or
Circuit card 2A1A10 or 2A1A16 faulty ........ Replace circuit card 2A1A10 (para 4-18) If
LEVEL indicator does not light.
this fails to correct the trouble, replace
56
Counter indication is not as spe-
Circuit card 2A1A15 or 2A1A10 faulty.
Replace circuit card 2A1A15 (para 4-
cified.
18) If this fails to correct the trouble,
57
Scope indication is not as spe-
Circuit card 2A1A2 faulty.
cified.
58
Scope indication is not as spe-
Circuit card 2A1A3 faulty.
cified.
59
LEVEL Indicator does not go out
Circuit card 2A1A10 or 2A1A16 faulty
or cage indicator does not light.
If this does not correct the trouble, replace
60
Scope Indication is not as
Circuit card 2A1A4 faulty.
Replace circuit card 2A1A4 (para (4-18).
Indicated.
Change 1 4-7