TABLE 7-2. CONSOLE TEST PROCEDURES (cont)
Normal
Circuit or
If Indication
Procedure
Step
Indication
Function Tested
Is Abnormal
READOUT SELECT to 0.
3
Part
E1 0 to 10 VOLTS of
Console to 0.
(cont)
E1 vernier of Console to 0.
S1 of Field Tester to OFF.
While observing multi-
meter, set E 1 5 to 90
VOLTS to each position
from 5 through 90.
E1 5 to 90 VOLTS to 5.
3
Same as part I,
Same as part I, this
Same as part I, this
Part
this step.
step.
step.
While observing multi-
II
meter, set E1 0 to 10
VOLTS to each position
from 1 through 10.
El 5 to 90 VOLTS to 5.
Same as part I, this
Same as part I, this
3
indicate 6.00
step.
step.
Part
0.30 volts.
III
E1 0 to 10 VOLTS to 0.
While observing multi-
meter, set E1 vernier to
500.
Ssme as part I, this
3
While observing multi-
Same as part I, this
indicate 7.00
step.
Part
meter, set E1 vernier to
step.
1000.
0.35 volts.
IV
E2 Voltage Range Test
4
Same as step 3, except
Same as step 3,
Same as step 3 except:
except applicable
applicable to E2 power
voltage supply output
to E2 VOLTS and
supply circuit.
at E2 switch 1S2A5
S2 of field tester to E2.
vernier controls.
and 8.
E2 NORM. -REV. of
Console to E2 NORM.
While observing multi-
meter, set E2 vernier as
in step 2, parts I, II, III,
and IV.
E1/E2 Load Regulation
5
Check
Replace subassemblies
This test checks regu-
Set up equipment as in
to isolate trouble to E1
lation of E1 and E2
indicate 0.050 volt
or E2 power supply
circuits by connecting
dc or less at each
circuit.
setting.
(Continued)
(Continued)
(Continued)
7-6